Manual Cleaning; Manual Disinfection - Mpv Medical Microdrop Rf6 Plus Gebrauchsanweisung

Professioneller vernebler für kinder und erwachsene
Vorschau ausblenden Andere Handbücher für MicroDrop RF6 plus:
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

EN

2. Manual Cleaning

An alternative method to mechanical cleaning and disinfection
is the manual method:
a. Equipment
Cleaning agent Korsolex® Endo Cleaner 0.5 % (Bode)
Brush
Warm water (approx. 50°C)
b. Procedure
Disassemble completely the nebulizer and place each part
and the accessories in the cleaning solution.
Clean the parts thoroughly with a brush.
Carefully rinse the components with hygienically tested
water and let them dry completely on a clean, dry and
absorbent surface.
c. Validation
Duration of cleaning procedure: approx. 5 minutes

3. Manual Disinfection

Select a suitable disinfectant solution that has been proven to
be effective and compatible with the materials of the accesso-
ries (PP , TPE, PVC, Silicone):
a. Equipment
Disinfectant Korsolex® Basic 4% (Bode)
Water
b. Procedure
Disassemble completely the nebulizer and place each part
and the accessories in the disinfectant solution.
Leave the accessories in the solution for the time indicated
in the instructions of the disinfectant
Carefully rinse the components with hygienically tested
water and let them dry completely on a clean, dry and
absorbent surface.
c. Validation
Treatment time: about 15 minutes
22
4. Sterilization
a. Equipment
Steam sterilizer – preferably with fractionated pre-vacuum
– in accordance with DIN EN 285 or DIN EN 13060 (type B)
b. Procedure
Disassembly the nebulizer as shown in pictures and pack
the components together with the accessories in a sterile
barrier system in accordance with DIN EN 11607 (e.g. foil-
paper packaging)
Maximum sterilization temperature and holding time:
121 °C, at least 20 min. or
134 °C, at least 3 min.
c. Validation
The sterilisation procedure is validated in its conformity to
ISO 17665-1.
NOTE: A validated sterilization process can only be carried out
on products that have been cleaned and disinfected.
LIMITS
The nebuliser can be sterilised up to 300 times and used for
up to one year. Do not expose the nebuliser or accessories to
temperatures above 134° C. The connection tubing cannot be
processed for re-use.
Use original manufacturer accessories and spare
parts only. The manufacturer will not be liable if
non original accessories or spare parts are used.