2. Disinfection

THERMAL DISINFECTION

Disinfectable accessories are: D1 – D6 (without rubber band),
E1 – E6
Method A (boiling water)
Disinfect the accessories by boiling in water for 20 minutes
(do not boil the tubing!) - preferably with distilled or demine-
ralized water to prevent limescale. Please add enough water
to the pot to prevent the nebulizer parts from melting by
touching the bottom of the pot. After boiling, shake out the
accessories and place them on a paper towel or kitchen towel,
or dry with a warm jet of air (such as a hairdryer). Note: Do
not boil the tube (Pic. D7)!
Method B (baby bottles vaporiser)
Disinfect the accessories with a steam sterilizer for baby bottles
(do not use microwave sterilizers). Perform the process
exactly as directed by the sterilizer manufacturer's manual.
To ensure that the disinfection is effective, choose a sterilizer
with a cycle of at least 6 minutes.
18
3. Chemical disinfection
Disinfectable accessories are: D1 – D6 (without rubber band),
E1 – E6
The disinfection procedure described in this section should
be performed before using the accessory. Disinfection is only
successful if every step is followed carefully and the compo-
nents have been cleaned beforehand.
Note:
a) Use a container of a suitable size to house all the compo-
nents to be disinfected. Fill it with a 2% -5% sodium hypo-
chlorite solution. Please strictly follow the instructions of
the disinfectant manufacturer, especially regarding safety
instructions, concentration, dosage and exposure time.
Ready-to-use solutions that you do not need to mix your-
self can be bought on the internet or in pharmacies.
b) Immerse each component completely in the solution. Make
sure that there are no air bubbles. These would prevent the
disinfectant coming into contact with the components. The
disinfection would not be completely possible.
c) Remove the disinfected components after the exposure
time and rinse thoroughly with warm drinking water. After
disinfecting, shake out the accessories and place them on
a paper towel or kitchen towel, or dry them with a warm
jet of air (such as a hairdryer). Dispose of the disinfectant
solution according to the manufacturer's instructions.