Ultimate Comfort
Guidance and manufacturers declaration - electromagnetic emission
The Dental Chair is intended for use in the electromagnetic environment specified below. The customer or the user of the model Dental Chair should assure that it
is used in such an environment.
Emissions test
RF emissions
according CISPR 11
RF emissions
according CISPR 11
Harmonic emissions
according IEC 61000-3-2
Voltage fluctuations/flicker emissions
according IEC 61000-3-3
Guidance and manufacturers declaration - electromagnetic immunity
The Dental Chair is intended for use in the electromagnetic environment specified below. The customer or the user of the Dental Chair should assure that it is used
in such an environment.
Immunity tests
Electrostatic discharge (ESD)
according IEC 61000-4-2
Radiated RF disturbance
according to IEC 61000-4-3
Electrical fast transient/burst
according IEC 61000-4-4
Surge
according IEC 61000-4-5
Interruptions and voltage varia-
tions on power supply input
lines
according to IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic filed
according IEC 61000-4-8
REMARK
U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration – electromagnetic immunity
The Dental Chair is intended for use in the electromagnetic environment specified below. The customer or the user of the Dental
Chair should assure that it is used in such an environment.
© Ritter Concept GmbH
C20-0087-1601
Compliance
Group 1
Class B
Class A
Compliant
IEC 60601-test level
Compliance level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3 V/m
3 V/m
80 MHz to 2.5 GHz
±2 kV
for power supply lines
±2 kV
±1 kV
for input/output lines
±1 kV
±1 kV line(s) to line(s)
±1 kV line(s) to line(s)
±2 kV line(s) to line(s)
±2 kV line(s) to line(s)
<5 % U
<5 % U
T
(>95 % dip in U
for 0.5 cycle)
(>95 % dip in U
T
40 % U
40 % U
T
(60 % dip in U
for 5 cycles)
(60 % dip in U
T
70 % U
70 % U
T
(30 % dip in U
for 25 cycles)
(30 % dip in U
T
<5 % U
<5 % U
T
(>95 % dip in U
for 5 s)
(>95 % dip in U
T
3 A/m
3 A/m
Electromagnetic environment - guidance
The unit uses high frequency energy exclusive for its internal function. From
there the high frequency emission is very low and it is improbable, that other
electronical devices are being disturbed.
The unit is determined for the usage in all facilities including living areas and
such, which are directly connected to a public power supply that supplies
also buildings which are used for living purposes.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
d= 1.17 · root(P/W) for 80 MHz to 800 MHz
d= 2.33 · root(P/W) for 80 MHz to 2.5 GHz
with P as the maximum of rated power of the transmitter
in watt according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
The field strength in each frequency of stationary radio
transmitters should be - in accordance with a site survey
- less than the compliance level.
In the vicinity of equipment that is marked with the logo
„Non-ionizing radiation", disruptions are possible.
for power supply lines
Mains power quality should be that of a typical commer-
for input/output lines
cial or hospital environment.
Mains power quality should be that of a typical commer-
cial or hospital environment.
Mains power quality should be that of a typical commer-
T
for 0.5 cycle)
cial or hospital environment. If the user of the Dental
T
Chair requires continued operation during power mains
interruptions, it is recommended that the Dental Chair be
T
for 5 cycles)
powered from an uninterruptible power supply or a
T
battery.
T
for 25 cycles)
T
T
for 5 s)
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.