Ultimate Comfort
Client (dental practice)
Device type:
ID No.:
Serial No.:
Safety inspection according: EN 62353 / VDE 0751
Visual inspection
Protective earth conductor not demaged (for insulation class 1 only)
Cover and mechanical parts not demaged
Insulating parts not demaged
Device connecting cables including plugs without defects
Labels exist or completed, nameplate correctly
Others
Measurement
Measurement
Protective earth resistance
Insulating resistance
Test voltage
Device leakage current (Equivalent device leakage current)
Encl. leakage current (Housing leakage current)
Leakage current of application part (Equivalent patient leakage current)
Test voltage leakage current (nominal Voltage ±10%)
Functional testing
Function and safety testing without defects
The unit has safety deficiencies:
Short-circuit L-N
Body circuit L/N-SL
Test cycle:
Next test date:
Used measurement device
Tester:
Place, Date
© Ritter Concept GmbH
C20-0087-1601
Inspection protocol
safety inspection according DIN EN 62353
...... months
...................
Type:
Model:
Signature
for
Contractor
Manufacturer:
Insulation class: 1
Type of applied parts: BF
Date of safety inspection:
measured value
Safety inspection
limit value
OK
<0.300 Ω
<10.000 mA
min.
......... V
max. ........ V
Anhang
Appendix
51