Eclipse 5
Alarms Table
Alarm
All OK, warm-up done
Warm-up O2 < 85%
No breath detected (less than
45 seconds)
No breath detected (exceeding
45 seconds)
One wire communication loss
Cannot charge battery
Ambient pressure fail
O2 < 85% (After warm-up)
Flow rate error > +/-10%
Warm battery > 60°C
Low battery
(5% remaining battery
capacity)
Loss of power
Malfunction
20 - ENG
PN 21182453-C4 E | User Manual
All manuals and user guides at all-guides.com
LCD Text
NO ALARM
O2 LOW
NO ALARM
P <--> C
ONE WIRE
CHARGER
AMB PRESS
O2 LOW
FLOWRATE
WARM BAT
LOW BAT
********
FAIL XX
Green
Yellow
Audio
ON
OFF
OFF
ON
ON
No Beeps
ON
OFF
OFF
ON
ON
1 alarm beep, repeat
every 20 seconds
ON OR
ON
1 alarm beep, repeat
OFF
every 20 seconds
ON OR
ON
1 alarm beep, repeat
OFF
every 20 seconds
ON OR
ON
1 alarm beep, repeat
OFF
every 20 seconds
OFF
ON
1 alarm beep, repeat
every 20 seconds
OFF
ON
1 alarm beep, repeat
every 20 seconds
ON
ON
1 alarm beep, repeat
every 20 seconds
ON
ON
1 alarm beep, repeat
every 20 seconds
OFF
ON
1 alarm beep, repeat
every 20 seconds for
200 seconds
OFF
ON
1 alarm beep, repeat
every 20 seconds for
200 seconds
Code
000
008
000
200
100
001
004
008
020
002
010
040
080