Keep the device safe from impacts and especially from dropping.
•
When removing the gutta percha injection needle and heat insulation protection cover, turn off the
power first in order to avoid any accidental starting and injury that might result from touching the trigger
accidentally.
•
When the battery power icon shows low power, please charge in time using the original power adapter.
•
Please use the original accessories, those from other manufacturers may not work with this equipment or
cause equipment damage.
•
Please do not disassemble or repair the equipment without permission, otherwise the warranty will no
longer be valid.
•
This equipment should not be placed in wet places or be in contact with any liquid.
•
Do not expose the equipment to a direct or indirect heat source. The equipment should be operated and
kept in a safe environment.
5.
Troubleshooting
Fault
The device cannot be
switched on
The gutta percha does not
flow out from the injection
needle
The gutta percha injection
needle cannot be removed
With the power adapter
plugged in, the device is not
charged
With the battery fully
charged, the service cycle
becomes shorter
6.
Cleaning, disinfection and sterilization
6.1 Cleaning of the charging base and the central unit
The charging base and the surface of the central unit can be wiped with a soft cloth dipped in a small amount
of neutral detergent or alcohol.
6.2 Cleaning and disinfection of the gutta percha injection needle, silver
needle bending device and heat insulation protection cover
Step
1. Rinse
2. Wipe
3. Brush wash
4. Soak
Possible cause
1. Insufficient battery power.
2. Short circuit of the charging
interface.
3. The hot melt gutta percha filling
device is damaged.
1. The gutta percha quantity progress
bar has reached the end, indicating
the gutta percha bar has been used
up.
2. The gutta percha injection needle is
damaged or clogged.
The melted gutta percha material
has cooled down in the cavity and
hardened
1. Not correctly plugged in.
2. Power supply is damaged or the
specifications do not match.
3. The charging base contacts are
dirty.
Reduced battery capacity.
Wash the gutta percha needle, silver needle bending device, heat insulation
protection cover with running water for 2 min to remove the surface contaminants.
Wet a soft clean cloth with a cleaning agent and thoroughly wipe the surface of
the product 5 times. Replace the soft cloth after each wipe. If there are still visible
pollutants remaining, wipe repeatedly until there are no visible pollutants.
Brush the product thoroughly with an instrument brush for 3 min.
Immerse the product in the cleaning agent for 5 min.
1. Charge the device connecting it
to a power supply.
2. Remove the foreign bodies at
the interface, wipe it clean and
reconnect it to the power supply.
3. Contact the local dealer or our
company.
1. Turn on the gutta percha load
function manually.
2. Replace the gutta percha
injection needle.
1. Turn on the power switch, press
the trigger, and pull the needle
out when the heating indicator
is on.
2. Contact the customer service.
1. Pull out and reconnect.
2. Contact your dealer.
3. Wipe the charging base
contacts with alcohol.
1. Send the device to the
maintenance department for
maintenance.
2. Contact the customer service.
Process
Solution
9