to the main interface) and press the trigger to return to the main interface. 3 RESET (quick reset). Press the
setting button to select RESET (quick reset). Press the trigger to start the quick reset program. After the quick
reset, the display screen displays RESET COMPLETE.
1 RELOAD GP (automatic residual gutta percha removal and reset). Short press the setting button to select
RELOAD GP (automatically remove the residual gutta percha and reset). Press the trigger to start the program
to automatically remove the remaining gutta percha and reset. After the display screen shows RELOAD GP, the
dental gutta percha bar can be installed as described in section 2.6.
2 EXIT (return to the main interface). Press the setting button to select EXIT (return to the main interface) and
press the trigger to return to the main interface.
3 RESET (quick reset). Press the setting button to select RESET (quick reset). Press the trigger to start the
quick reset program. After the quick reset, the display screen displays RESET COMPLETE.
•
2.5.3 Heating interface
Short press the trigger for heating, if you need to stop heating, short press the power button.
The contents displayed on the heating interface as shown in the Figure (a) are: heating state (HT), real-time
temperature and remaining gutta percha quantity progress bar. The indicator flashes when heating starts, and
it is always on after heating. After heating, the contents displayed on the heating interface as shown in Figure
(b) are: heating completion state (OK), preset temperature and remaining gutta percha quantity progress bar.
•
2.5.4 Push interface
After heating, long press the trigger to push the gutta percha. The gutta percha push interface, shown in the
figure, includes the remaining gutta percha quantity progress bar, the heating completion state (OK), the
preset temperature and the direction indication arrow. When the gutta percha bar is exhausted, reset starts
automatically.
2.6 Loading the gutta percha bar
Before loading, you need to confirm that the device is in the state of «reset complete» (in the case of newly
purchased products, the gutta percha bar can be directly loaded without using the «reset» state). The gutta
percha bar must first be inserted into the gutta percha injection needle, then attach the gutta percha injection
needle to the device as described in section 2.3.
Attention :
Before loading the gutta percha bar, please confirm that the gutta percha remaining in the heating cavity
has been removed. If not, please perform the RELOAD GP (automatic removal and reset) procedure on the
operation interface (refer to 2.5.2 for details) before loading. If the device is automatically reset, the gutta percha
can be loaded.