If the weight applied to the platform exceeds the maximum capacity
of the REFSCALE-C (100 kg), the display will show "OOR", which
means "Out of Range". This applies to the absolute weight
measurement.
To avoid damage to the REFSCALE-C, remove the weight
from the platform immediately.
9.2

Troubleshooting

Malfunction
Device cannot be switched on
OOR
Stable measured value is not
met
Display does not change
Wireless connection not
working
ZERO does not work
Error Display «
27
-.--
»
Solution
Battery inserted incorrectly.
Battery contact break was
not pulled out.
Weight over 100 kg.
Reduce weight immediately.
Re-align the device.
Restart the device.
Hold function is activated
(Hold function symbol is
displayed).
A device is already
connected. This can be seen
when the Wireless
connection symbol lights up
continuously.
Weight not close to zero.
Remove weights from the
device.
Restart the device.
Remove weights from the
device and realign it.
Press ZERO to zero the
display.
Restart the device.