a) Removing pump head
u Mark the position of head plate
, cover
and cover plate
to each other by a drawing line with
a pencil. This helps avoid incorrect
assembly later.
v Undo the 4 screws
plate and lift the head plate with the
cover
off the pump housing.
w Mark the position of intermediate
plate
and housing
each other by a drawing line with
a pencil.
x Lift the intermediate plate
housing .
b) Change structured diaphragm
u Using a small screwdriver, between
the housing
and the outer edge
of the structured diaphragm ,
carefully lever the edge of the dia-
phragm lightly upwards.
v Grip the structured diaphragm
opposite sides, unscrew it about two
turns (anti-clockwise).
w Hold the pump with one hand,
so that the head is pointing
downwards. Turn the structured
diaphragm
anti-clockwise to
unscrew it.
x Take the diaphragm support
and diaphragm spacer(s) k off the
threaded portion of the diaphragm
and retain them.
y Check that all parts are free from
dirt and clean them if necessary
(see section 6. Cleaning).
z Put the diaphragm support
diaphragm spacer(s) , in that order,
on the threaded portion of the new
structured diaphragm .
{ Screw the new structured diaphragm
, complete with diaphragm sup-
port
and diaphragm spacer(s)
into the connecting rod (clockwise)
and tighten it by hand.
6
c) Changing the valve plates
u Remove the valve plates
relative
ing rings
plate
cling to the head plate).
v Check that the valve seats in the
in the head
head plate
tion are evident on these
parts they should be replaced.
relative to
w Lay the new valve plates
recesses in the intermediate plate
pressure sides are identical, as are
off the
upper and lower sides of the plates.
x Check that the valve plates
not deformed by moving them
gently sideways in their recesses.
y Lay the new sealing rings
the intermediate plate .
d) Refitting the pump head
on
u Place the intermediate plate , with
valve plates
on the housing , in the position
indicated by the drawing line.
v Place the head plate , with cover
indicated by the drawing line.
w Check that the head plate
is centred by moving it gently
sideways.
x Tighten the screws , evenly and
diagonally, first gently, then firmly.
• If you have any questions about ser-
vicing call our technical adviser (see
last page for contact address).
and
and seal-
from the intermediate
(sealing rings could
and intermediate plate
are clean. If scratches or distor-
in the
. The valve plates for suction and
are
on
and sealing rings
on the housing , in the position
6
4 3
5
Fig. 4: Pump parts
Fig. 5: Pump parts
Specification
Housing
Intermediate plate
Head plate
Screw
Cover
Cover plate
Valve plate
Sealing ring
Structured diaphragm
Diaphragm support
Diaphragm spacer(s)
Connecting rod
Eccentric
Counter weight