Periodic errors.
1. The chair won't charge every a empt.
2. The chair will only work periodically, even a er a reset.
The chair is eleva ng uneven, one defect
1. If one of the actuators on the chair isn't eleva ng and
the chair is being pulled sideways.
HMN a/s
If experiencing periodic errors always try and reset the
chair as described above.
It is also possible that the issue is an incorrectly mounted
wire, so check the wires is plugged correctly in.
If the problem con nues it might be a defect switch box.
Contact HMN for further advice.
If you o en experience issues with the chair not working
and the resets only work temporarily before all lights in the
hand control are blinking again it might be sign of incor-
rectly mounted wires or issues with the switch box. Contact
HMN for further advice.
The actuators on the chair will not always go at the same
speed. This might result in slighty lted seat frame. In this
case just lower the chair completely to reset the pistons in
the actuators.
If there is a problem with one actuator not working it might
pull the chair sideways, and this indicates there could be a
problem with that actuator.
Do NOT try to elevate the chair if this is the case.
With this issue we need to know if the frame has been
pulled out of shape. Send us pictures of the frame from the
front and back, and we will be able to tell if the frame is
damaged.
If not, just changing the actuator is enough, but if there is
damage to the frame, we might need the chair send to us
for repair, or will send you the needed parts to replace.
With these issues always contact HMN.
www.HMN.dk
28