Instrument not
ERR 003
responding
No tone, no light
ACC
when placing the
®
HandyStep
electronic
into the charging dock
Charge-indicator is
No message
flashing
Dispensing process
ERR VOL
interrupted.
Attention!
Volume not completely
discharged
Note:
If other error messages are displayed, remove battery pack and insert it again.
If this does not solve the problem, send instrument in for repair.
Electronic fault or tip connector
not at lowest position when
inserting new tip
- Instrument contains no battery
- Battery pack inserted improperly
- Battery pack faulty
Removal or insertion of a new
battery pack
Dispensing process interrupted by
- Pressing the ∆ key during
dispensing
- electronic faults
- 37 -
®
4. HandyStep
electronic automatically carries out a
reference run.
5. Display: no TIP
Insert new tip (see page 16).
1. Press ENTER to confirm error.
2. Follow MENU-instructions:
- Display rEF (reference run)
- Press STEP-Key
Attention: Liquid remaining in tip will be discharged
during reference run (rEF). Discharge sample into
suitable vessel (see page 5, point 10).
If error message is still shown on the display remove
battery pack and insert it again (reset). If this does not solve
the problem send instrument back for repair.
- Insert battery pack.
- Make sure battery contacts are aligned with contacts
in the instrument.
- Replace battery pack.
Charge battery pack for at least 3.5 hours inside
®
the HandyStep
electronic.
1. Press ENTER to confirm error.
2. Restart dispensing by pressing the STEP key.