LASER LEVEL
Check of vertical line
Use a plumb bob as a reference of a vertical. Place the instrument at
distance of 1,5 m from the plumb bob.
1. Move lever lock to unlocked position, then switch on vertical
plane and align vertical laser line with the low point of plumb bob.
2. If deviation between the laser line and the plumb line exceeds
±0,3 mm per 1 m of the plumb bob length (for a 2,5 m plumb bob
deviation shouldn't exceed 0,75 mm) please contact service center.
14
14
CONDTROL NEO X200
user manual