Adjustment
If incorrect reference values are used in the adjustment, incorrect
measurement values will be displayed later.
The adjustment can be interrupted at any time.
Information from last adjustment
Click on Get Last Adjustment info.
- The values are shown on the screen and PC con appears in the display
of the instrument for about 2 sec.
Pressure (testo 622 only)
Click on Adjust Pressure and confirm message with Yes.
Follow the instructions and enter values.
1
Click on Adjust Pressure.
2
- The values are calculated and shown on the screen and PC con appears
in the display of the instrument.
Confirm message with Yes and return to the adjustment menu with Exit.
3
Cancel adjustment with Exit.
Humidity
The equalization period for a humidity adjustment should be at least 30 min.
The entered equalization period for the adjustment is valid with an inflow
> 0.5 m/s. With a lower inflow in the air conditioning chamber, the
equalization period is to be extended sensibly until the system is balanced
out.
Click on Adjust Humidity and confirm message with Yes.
Follow the instructions and enter values.
1
Click on Adjust Humidity.
2
- The values are shown on the screen and PC con appears in the display
of the instrument.
Confirm message with Yes and return to the adjustment menu with Exit.
3
Cancel adjustment with Exit.
Temperature
The equalization period for a temperature adjustment should be at least 2 to
3 h.
The entered equalization period for the adjustment is valid with an inflow
> 0.5 m/s. With a lower inflow in the air conditioning chamber, the
equalization period is to be extended sensibly until the system is balanced
out.
Using the product 19