The Theragun SENSE is intended for use in the electromagnetic environment specified below. The customer or the user of the Theragun SENSE should assure that it is used in such an
Immunity test
Voltage dips, short interruptions and voltage variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Guidance and manufacturer´s declaration – electromagnetic immunity
environment.
IEC 60601 test level
0 % UT; 0,5 cycle g) At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30 cycles Single phase: at 0°
0 % UT; 250/300 cycle
30 A/m
NOTE
UT is the a. c. mains voltage prior to application of the test level.
Compliance
Electromagnetic environment -guidance
level
Mains power quality should be that of a typical
commercial or home environment. If the user
of the Theragun SENSE requires continued
N/A
operation during power mains interruptions, it
is recommended that the Theragun SENSE be
powered from an uninterruptible power supply
or a battery.
Power frequency magnetic fields should be
30 A/m
at levels characteristic of a typical location in a
typical commercial or home environment.
EN
11