Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
EN
The Theragun SENSE is intended for use in the electromagnetic environment specified below. The customer or the user of the Theragun SENSE should assure that it is used in such an
Immunity test
Electrostaticdischarge (ESD)
IEC 61000-4-2
Electrostatic transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
10
Guidance and manufacturer´s declaration – electromagnetic immunity
environment.
IEC 60601 test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
Electromagnetic environment
Compliance level
-guidance
± 8 kV contact
Floors should be wood, concrete
or ceramic tile. If floors are covered
± 2 kV, ± 4 kV, ± 8 kV,
with synthetic material, the relative
± 15 kV air
humidity should be at least 30 %.
Mains power quality should be that of a
N/A
typical commercial or home environment.
Mains power quality should be that of a
N/A
typical commercial or home environment.