CATAPULT 5000 SERIES SCALES
6.
MAINTENANCE
CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.
6.1
Cleaning
• The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
• Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panel.
6.2
Troubleshooting
Error Code
Error 8.1
Pow er On Error
Error 8.2
Pow er On Error
Error 8.3
Over Range Error
Error 8.4
Under Range Error
Error 8.5
Tare out of range Error
Error 8.6
Display Overflow
-------
Busy message
--NO--
Action not allow ed message
Battery icon flashing
Low Battery error
CAL E
Calibration Error
Lo.rEF
Low reference w eight w arning
message
rEF.Err
Unacceptable reference
w eight message
6.3
Service Information
If the troubleshooting section does not resolve your problem, contact an authorized Ohaus Service Agent. For
Service assistance in the United States, call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM Eastern
Standard Time. An Ohaus Product Service Specialist will be available to assist you. Outside the USA, please
visit our website www.ohaus.com to locate the Ohaus office nearest you.
7.
TECHNICAL DATA
7.1
Specifications
The technical data is valid under the following ambient conditions:
Indoor use only
Ambient temperature:
Relative humidity:
Altitude:
Mains supply voltage fluctuations: up to ± 10% of the nominal voltage
Installation category:
Pollution degree:
Protection class:
TABLE 7-1. TROUBLESHOOTING.
Description
-10° C to 40° C / 14° F to 104° F
Maximum relative humidity 80% for temperatures up to 31° C decreasing
linearly to 50% relative humidity at 40° C.
up to 2000m
II
2
III
Weight reading exceeds Pow er On Zero limit.
Weight reading below Pow er On Zero limit.
Weight reading exceeds Overload limit.
Weight reading below Underload limit.
Tared at one unit but after sw itching to another unit the tare
value exceed the max.
Weight exceeds 6 digits. Happened in the cases of
Accumulation or counting PCS display
Displayed during tare setting, zero setting, printing
Function not executed.
Battery is empty
Calibration value outside allow able limits
Average Piece Weight too small. (Warning)
Reference Weight too small. The w eight on the pan is too small
to define a valid reference w eight.