6 Starting Up
6.1 Installation Site Requirements
CAUTION:
Measurement errors and damage to the pyrometer through
• Ambient temperatures too high
• Strong contamination of the optics due to dust, smoke, steam or other causes like air pollution
• Electromagnetic interference sources
You must take into account the following climatic conditions and the requirements of the place of use
of the pyrometer!
6.1.1 Climatic conditions
The following requirements must be fulfilled with respect to the climatic conditions at the place of use of the
pyrometer:
• Temperature range 0 °C . . . +40 °C (up to +80°C with connected cooling)
• Maximum air humidity: 60 % at room temperature
• Atmosphere free of dust, corrosive vapours, and fume gases
6.1.2 Requirements at the place of use
Take into account the following requirements at the place of use of the pyrometer:
• Take care with the choice of the place of use and take into account the ergonomic and the legal guidelines
for industrial safety in order to ensure safe operation of the pyrometer.
• Set up the pyrometer on a firm, stable base. The base must be free from concussion and vibration.
Note:
We recommend using the rail mounting plate available as an accessory for the attachment of the pyrom-
eter as well as a clamping attachment (see Accessories (section 3.3) at page 4).
• Do not expose the system to direct sunlight or radiation from heating (e.g. radiators).
• Do not set the pyrometer near strong electromagnetic fields or near electromagnetic interference sources.
• Make sure that the pilot light falls unhindered on the object being measured and that there is nothing in
the path of the rays.
Take into account the following requirements at the place of use of the bench unit:
• Place the bench unit on a preferably flat, big enough surface.
• The maximum distance from the bench unit to the PC is 5 m due to maximum USB cable length specifi-
cations.
• The distance between pyrometer and bench unit should not exceed 20 metres.