„ 7.1. Battery Replacement
The battery compartment is located at the bottom of CanalPro Apex Locator and its cover is secured by a
screw.
7.1.1.
Release the screw and remove the battery compartment cover (Pic. 19).
7.1.2.
Remove the battery from battery compartment and disconnect the battery cable jack from the
battery connector of CanalPro Apex Locator (Pic. 20).
7.1.3.
Insert the cable jack of the new battery into the battery connector.
7.1.4.
Insert the battery into the battery compartment.
7.1.5.
Close the battery compartment and secure the cover with the screw.
Picture 19
8. Maintenance, cleaning and sterilization
„ 8.1. General
• The device does not contain user serviceable parts. The service and repair should be provided by
factory trained service personnel only.
• All objects that were in contact with potentially infectious agents should be cleaned after each use:
Lip Clip, File Clip and Touch Probe should be disinfected and sterilized by autoclaving between tre-
atments. Please follow "Disinfection and sterilization procedure" described in section 8.2.
Measurement cable, the device and its cradle should be cleaned using tissue or soft cloth impreg-
nated with aldehyde free disinfecting and detergent solution (a bactericidal and fungicidal).
The measuring cable cannot be autoclaved.
Use of agents other than specified above may cause damage to the equipment and its accessories