18.7.2 Perform internal adjustment
The balance must not be switched off, disconnected from the power
supply or loaded during internal adjustment
18.8 Output of the adjustment test result
The scale provides the option to automatically show the result on the interface after
adjustment or after the adjustment test.
To use this function, you must activate <41 RS232 C> (see chapter 21.7.1).
18.9 Automatic power on when connected to mains
If the automatic power-on function is enabled, the scale will switch on automatically
when it is connected to the mains. Users then no longer need to press the [ON/OFF]
key. This function can be used, for example, when the scale is used in conjunction with
other equipment.
70
 Switch on balance
If internal adjustment is activated every time the
balance is switched on:
 The scales perform the internal adjustment and
automatically switch to measuring mode.
If internal adjustment is activated on request:
 Select the desired setting with [F3] and [F4].
YES
Perform internal adjustment
NO
Do not perform internal adjustment
 The scales perform the internal adjustment and
automatically switch to measuring mode.
 Open <64 SCALE MANAGE> (see chapter 18.1).
 Use the [F1] and [F2] keys to navigate to <645 SPAN
OUT>.
 Press [F4] key to make settings
 Use keys [F1] and [F2] to select the desired setting:
OFF
Disabled
ON
Enabled
 Press [F4] key to confirm selection.