Herunterladen Diese Seite drucken

PCE Instruments PCE-VT 3800S Bedienungsanleitung Seite 46

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

As described previously, the ISO evaluation function requires the measuring unit velocity or
displacement in conjunction with the parameter RMS. Otherwise, the menu to enable this function
cannot be opened and the following hint is displayed on the screen.
Machine groups:
Group 1:
Group 2:
Fig. 6 Hint ISO evaluation disabled
Fig. 7 Hint for activation of ISO evaluation
Large machines with nominal power >300 kW;
Electrical machines with shaft height >315 mm
These machines generally have plain bearings and the relatively
high
rated/operating
−1
15,000 ������
.
Medium-sized machines with nominal power between 15 kW and
300 kW;
Electrical machines with shaft height 160 mm <H <315 mm.
© PCE Instruments
41
speeds
range
from
−1
120 ������
to

Werbung

loading

Diese Anleitung auch für:

Pce-vt 3900Pce-vt 3800