Herunterladen Inhalt Inhalt Diese Seite drucken

OHAUS Valor V31XH2 02 Bedienungsanleitung Seite 7

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 51
Valor 3000 Series Scales
3.2 Percent Mode
Use this mode to measure the weight of a sample as a percentage of a reference
weight.
1. Press MODE/UNIT/Menu until Set.rEF and the % icon are displayed.
2. Press No to use the reference weight in memory and proceed to step 6.
3. Press Yes to establish a new reference. Scale display will now blink until
reference is stored.
4. Put an empty container on the pan, then press O/T/On/Off to tare the container
weight.
5. Add the desired amount of material to the container, then press ENTER/Cal to
store the displayed weight. The display shows 100%.
6. Replace the reference material with new material. The display shows the
percentage of the sample compared to reference weight.
3.3 Specific Gravity Mode
Use this mode to measure the specific gravity of a sample.
1. Press MODE/UNIT/Menu until Set.rEF and the S icon are displayed.
2. Press No to use the reference weight in memory and proceed to step 6.
3. Press Yes to establish a new reference weight. Scale display will now blink until
reference is stored.
4. Put an empty container on the pan, then press O/T/On/Off to tare the
container weight.
5. Fill the container with water, then press ENTER/Cal to store the displayed
weight. The display shows the water weight with the S icon blinking.
6. Replace the water with the same volume of sample material, then press
ENTER/Cal. The display shows the specific gravity of the sample.
3.4 Fluid Ounce Mode
Use this mode to measure the fluid ounce volume of a sample.
1. Press MODE/UNIT/Menu until the FL icon is displayed.
2. Put an empty container on the pan, then press O/T/On/Off to tare the
container weight.
3. Fill the container with sample material. The display shows the volume of the
sample.
EN-5

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis