Herunterladen Inhalt Inhalt Diese Seite drucken

KERN RFC Serie Betriebsanleitung Seite 59

Preisrechnende waage
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization. Ensure that there are no objects on the weighing pan.
Arrange adjustment weight, details see chap.1 „Techn. data"
To turn on/off operate the on/off switch at the
underneath side on the right of the scale and
keep pressed for a little while. During the selftest
press
and keep pressed until Pn is
displayed.
Use the numeric keys to enter the standard
password 0000. Pn---- is displayed.
Confirm password with
, the adjustment
function F0 CAL will be displayed.
Press adjustment switch on the lower side of the balance.
Press
, UnLoAd will be displayed. Ensure
that there are no objects on the weighing pan.
Press
again, CAP LOADxx will be
displayed.
(example)
RFC-BA-e-2113
19

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis