Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 35
Hand wash all surfaces of the stretcher with warm water and mild detergent. Dry thoroughly. DO NOT
STEAM CLEAN, PRESSURE WASH, HOSE OFF OR ULTRASONICALLY CLEAN. Using these methods
of cleaning is not recommended and may void this product's warranty.
Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow disinfectant to evaporate. (Ap-
propriate disinfectant for nylon Velcro should be determined by the hospital.)
In general, when used in those concentrations recommended by the manufacturer, either phenolic type or
quaternary type disinfectants can be used. Iodophor type disinfectants are not recommended for use be-
cause staining may result. The following products have been tested and have been found not to have a harm-
ful effect WHEN USED IN ACCORDANCE WITH MANUFACTURERS RECOMMENDED DILUTION.*
TRADE NAME
DISINFECTANT
A33
Quaternary
A33 (dry)
Quaternary
Beaucoup
Phenolic
Blue Chip
Quaternary
Elimstaph
Quaternary
Franklin
Phenolic
Phenomysan
F2500
Franklin Sentinel
Quaternary
Galahad
Phenolic
Hi−Tor
Quaternary
LPH
Phenolic
Matar
Phenolic
Omega
Quaternary
Quanto
Quaternary
Sanikleen
Quaternary
Sanimaster II
Quaternary
Vesphene
Phenolic
Quaternary Germicidal Disinfectants, used as directed, and/or Chlorine Bleach products, typically 5.25% So-
dium Hypochlorite in dilutions ranging between 1 part bleach to 100 parts water, and 2 parts bleach to 100
parts water are not considered mild detergents. These products are corrosive in nature and may cause dam-
age to your stretcher if used improperly. If these types of products are used to clean Stryker patient handling
equipment, measures must be taken to insure the stretchers are rinsed with clean water and thoroughly dried
following cleaning. Failure to properly rinse and dry the stretchers will leave a corrosive residue on the surface
of the stretcher, possibly causing premature corrosion of critical components. Failure to follow the above
directions when using these types of cleaners may void this product's warranty.
REMOVAL OF IODINE COMPOUNDS
This solution may be used to remove iodine stains from mattress cover and foam footrest pad surfaces.
1. Use a solution of 1−2 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area.
Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to
soak or stand on the surface.
2. Rinse surfaces which have been exposed to the solution in clear water before returning bed to service.

Cleaning

MANUFACTURER
TYPE
Airwick (Professional Products Division)
Airwick (Professional Products Division)
Huntington Laboratories
S.C. Johnson
Walter G. Legge
Purex Corporation
Purex Corporation
Puritan Churchill Chemical Company
Huntington Laboratories
Vestal Laboratories
Huntington Laboratories
Airwick (Professional Products Division)
Huntington Laboratories
West Chemical Products
Service Master
Vestal Laboratories
1−13
*MANUFACTURER'S
RECOMMENDED
DILUTION
2 ounces/gallon
1/2 ounce/gallon
1 ounce/gallon
2 ounces/gallon
1 ounce/gallon
1 1/4 ounce/gallon
2 ounces/gallon
1 ounce/gallon
1/2 ounce/gallon
1/2 ounce/gallon
1/2 ounce/gallon
1/2 ounce/gallon
1 ounce/gallon
2 ounces/ gallon
1 ounce/gallon
1 1/4 ounce/ gallon

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Eye stretcher chair50505051

Inhaltsverzeichnis