Herunterladen Inhalt Inhalt Diese Seite drucken

YOKOGAWA FLXA402 Bedienungsanleitung Seite 7

Vorschau ausblenden Andere Handbücher für FLXA402:
Inhaltsverzeichnis

Werbung

Yokogawa Electric Corporation
Control drawing
Title
NFM038-A81
No.
Spec. restriction
Temp.
Power supply
class
AC version
100-240V AC
T4
DC version
12-24V DC
AC version
100-120V AC
T6
DC version
N/A
Notes:
1. No revision to this drawing without prior approval of FM.
2. Installation must be in accordance with the National Electrical Code (NFPA 70), ANSI/ISA-
RP12.06.01 and relevant local codes.
3. When installed in Division 2, Sensor 1 and Sensor 2 may be simple apparatus or nonincendive
field wiring apparatus meeting the conditions below, or alternatively, they may be equipment
suitable for Division 2 respectively, if a suitable wiring method other than nonincendive field
wiring is employed.
Ui ≥ Uo
Ii ≥ Io
Pi ≥ Po
Ci ≤ Co − Ccable
Li ≤ Lo − Lcable
4. WARNING-EXPLOSION HAZAERD. DO NOT OPEN WHILE THE EQUIPMENT IS
ENERGIZED OR WHEN AN EXPLOSIVE ATMOSPHERE IS PRESENT.
5. WARNING-EXPLOSION HAZAERD. DO NOT REMOVE OR REPLACE THE FUSE WHILE
THE EQUIPMENT IS ENERGIZED OR WHEN AN EXPLOSIVE ATMOSPHERE IS
PRESENT.
6. WARNING-SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR
DIVISION 2.
Model
02
Page
Revision
Number of
Ambient Temperature
mA
Output
-20 to
+55°C
2
4
2
4
2
4
FLXA402
0
2018-06-29
Date
Switch capacity
of Contact Output
-20 to
MAX. 5A
+40°C
IM 12A01F01-02DE-E
vi
MAX. 4A
8. Ausgabe: 25. Nov. 2022-00