Models FOB-NL:
Balance display
(example)
(example)
12
Operation
Switch off the balance
Press and hold
time, release
Press
, "0" is displayed
Put adjustment weight in the centre of the weighing
plate, wait for stability display and press
"CAL" is displayed, followed by the value of the
adjustment weight.
Take away adjustment weight
The balance changes into weighing mode.
Adjustment has now been completed successfully.
, press
3x at the same
- an internal value is displayed.
FOB-N-BA-e-1611
,