Operation

3.3

Intended use

The PID 5000 must only be used in connection with the Dräger
Air Guard. The PID 5000 is used to continuously measure
volatile organic compounds and is intended for use indoors.
The PID 5000 may only be installed and commissioned by
Dräger Service or authorised Dräger specialist personnel.
3.4

Limitations on intended use

Any use other than that mentioned in the "Intended use"
chapter shall be regarded as improper usage. The following in
particular is not permitted:
Use with devices other than the Dräger Air Guard.
Use for substances other than volatile organic compounds
of the approved oil types (see chapter 4.3.1 on page 21).
Measurement without gas treatment.
Measurement of air, which does not correspond to
EN 12021 and whose dew point is higher than -20 °C. This
means that the compliant condition must immediately be
restored if the limit value is exceeded.
20
4
Operation
WARNING
Danger due to incorrect setting
Any change to the PID 5000's settings may distort the
measurement result.
► Do not change the settings for:
Calibration
Current loop
Measuring time
Measuring range
Flow calibration
Operating hours counter lamp
► Only change settings that are described in the following
chapters.
4.1

Prerequisites

Perform a visual inspection for damage prior to every system
measurement.
WARNING
Danger due to devices that are not ready for operation
The first measured value is output 3 minutes after the
PID 5000 has been started.
► Take account of the start phase and wait until the
PID 5000 is ready for operation.