Description
Explanation of the status LED
Indicator
Explanation
Rechargeable battery is inserted and
fully charged (standby mode)
Status LED is
green.
Rechargeable battery is inserted and
being charged.
Status LED is flash-
ing green.
Rechargeable battery is not inserted.
Status LED is red.
Malfunction
Status LED is flash-
ing red.
When the rechargeable battery is fully charged, the charger
switches automatically to standby. In standby mode, the re-
chargeable battery stays fully charged at all times. In this mode
the rechargeable battery is neither overcharged nor damaged.
2.3

Functional description

The powered air purifying respirator is a respiratory protective
device depending on circulating air.
It filters the ambient air and makes it available as breathable
air. The device continuously takes in ambient air through the fil-
ter. The filter absorbs harmful substances depending on the fil-
ter type. In this way, the ambient air is recycled and finally
reaches the facepiece. There it is available as breathable air.
A continuous overpressure in the facepiece prevents ambient
air from penetrating.
2.3.1

Warning devices

Malfunctions during operation are indicated by warning devic-
es.
The warning devices include:
optical alarm (display on control panel)
acoustic alarm
vibration alarm
NOTICE
i
i
The vibration alarm is triggered in addition to the
acoustic alarm. Depending on the thickness and mate-
rial of the clothing, the vibration alarm might not be per-
ceived.
The fan unit always delivers the same default flow rate.
If in the foreseeable future the fan unit will no longer be able to
deliver the default flow rate (e.g. due to increasing saturation
of the particle filter), a warning or alarm is triggered.
18
2.4

Intended use

Depending on the employed filter type, the device protects
against particles, gases and vapours or combinations hereof.
2.5

Limitations on use

The device is not suitable for use:
in explosion-hazard areas (Ex-areas)
when there is a suspicion of contaminants with low warning
properties (smell, taste, irritation of eyes and airways)
in unventilated tanks, pits, canals etc.
2.6

Approvals

The device is approved according to
EN 12941:2009-02
EN 12942:2009-02
Therefore the device complies with directive 89/686/EEC on
personal protective equipment.
Additional directives as part of the CE marking:
EMC directive (2004/108/EC)
R&TTE directive (1999/5/EC)
NSR directive (2006/95/EC)
RoHS directive (2011/65/EU)
2.7
Explanation of type-identifying marking
and symbols
2.7.1

Name plates

Illustration of name plates on page 4
Fan unit:
Figure H
Standard battery
Figure I
charger:
Rechargeable bat-
Figure J
tery:
1 Product name
2 International Protection Code
3 Fulfilled EN standards
4 Symbol "Follow instructions for use"
5 WEEE symbol "Separate collection of electrical and elec-
tronic equipment"
6 Country of production
7 Manufacturer
8 CE marking
9 DataMatrix code with part and serial number
10 Serial number
11 Article code
12 Only for indoor use, not for outdoor use
13 Maximum ambient temperature
14 Electrical data
15 Pin assignment
16 Recycling symbol
17 Warning notice
Dräger X-plore 8000