Remote Alarm Contacts

All ES Series laboratory refrigerators include remote alarm contacts and these terminal contacts enable the user
to connect to a remote- c o n t r o l station; these contacts can be found on the rear of the cabinet and are
different for 158L and 288L as shown.
Remote Alarm Principle of operation
The remote alarm function of this appliance has been designed to go "OPEN CIRUIT" in the event of any of the
alarm conditions listed in the table below
Remote Alarm Contact Conditions (Default)
Alarm Condition
Normal condition (No Alarms)
High or low air temperature alarm
High or low simulated load temperature
Door open
Power failure
Table 10. Remote Alarm Contacts
158L Remote alarm contact -
288L Remote alarm contact -
CAUTION: In the case where the facility needs the contacts to close for an alarm condition, refer to
"Changing the Remote Alarm Contacts" section for reversing the relay condition.
ES Series Refrigerator
Remote Alarm Contact Data
Appliance as delivered
Max switching voltage
Max switching current
WARNING: These contacts must only be accessed by a competent electrician.
Relay Status Continuity
Closed
Open
Open
Open
Open
Voltage Free
240VAC
28VDC
2A
2A
Insert remote access
wires into red/black
terminal by depressing
lever, insert and release.
Insert remote access
wires into terminal by
depressing with small flat
driver, into positions 1 & 3
as shown
Yes
No
No
No
No