Herunterladen Inhalt Inhalt Diese Seite drucken

Changing The Remote Alarm Contacts; Lock/Unlock Controller - Thermo Fisher Scientific ES Serie Bedienungsanleitung

Labor-kühlschrank
Vorschau ausblenden Andere Handbücher für ES Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 30
Changing the Remote Alarm Contact
Follow this to change the alarm relay contact type.
There are two types – NC (normally closed) or NO (normally open)
Function
Press
Example screen
To change the alarm contact state
Scroll through to screen required.
(press button twice)
To display the current condition of remote
alarms (default will be NO, normally open)
To change to NC, normally closed
To confirm change to NC
To return to temperature display
Press for 1 second
Table 7. Remote Alarm Contacts
Lock / Unlock Controller
To lock the control panel, press the
buttons simultaneously on the controller for 5 seconds (display
shows "LOC"). Once completed the buttons can still be pressed but they will have no effect.
buttons simultaneously on the controller for 5 seconds (display shows "UNL")
To unlock press the
and the buttons can now be operated as normal.
13
ES Series Lab Refrigerator

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis