Installation and Operation

3.4 Checking the weighing platform

Function check
Before starting a weighing series, perform a function check of the weighing platform and
the connected weighing terminal.
1. Make sure that the weighing platform is connected to a weighing terminal and that the
weighing terminal is switched on.
2. Make sure that the weighing platform is unloaded and the display of the weighing
terminal shows 0.
3. Load the weighing platform.
The display must show a value different from 0.
4. Unload the weighing platform.
The display must return to 0.
Verification test
For a verification test refer to the User manual of the connected weighing terminal.
If the verification seal is broken, verification is no longer valid.

3.5 Important notes

For best weighing results observe the following:
▲ To achieve best weighing results, always place the weighing sample on the same
position on the scale.
▲ Avoid abrasive and wear processes.
▲ Avoid falling loads, shocks and lateral impacts.

3.6 Installation, configuration, service and repair

➜ For installation, configuration, service and repair of the weighing platforms call the
METTLER TOLEDO Service.