MD 9273 Leakage Clamp
Active power, reactive power and apparent power
Unit
W, VAr, VA
kW, kVAr, kVA
PF
Phase °
Note: One decimal place can be subtracted from the resolution following the rule:
Value of S checked in counts (disregarding the decimal point).
•
If it is above 2000 one decimal is subtracted from either P (where PF < 0.707) or Q (where PF >
•
0.707)
If it is below the decimal places remain as in the table.
•
Note: Max display resolution is 4 digits. Specified resolution may vary if result of the S value is longer than 4
digits.
The P and Q value show the same number of decimal places as the S value.
Range
1.600
16.00
24.00
160.0
240.0
1600
2.400
4.000
16.00
24.0
40.00
60.0
0.00 ... 1.00
-180.0 ... 180.0
Resolution
0.001
0.001
0.01
0.01
0.1
1
0.001
0.001
0.01
0.01
0.01
0.1
0.01
0.1
27
Technical Specifications
Accuracy
1 % ± 50 digits
± 5 digits
± 30 digits