Herunterladen Inhalt Inhalt Diese Seite drucken

CeraGem CGM MB-1101 Bedienungsanleitung Seite 48

Vorschau ausblenden Andere Handbücher für CGM MB-1101:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

2. Attached documents
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the
The CERAGEM-CH is intended for use in an electromagnetic environment in which
CGM MB-1101
radiated RF disturbances are controlled. The user of the CERAGEM-CH can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power (W) of
transmitter
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
46
CGM MB-1101.
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz
0.12
0.37
1.17
3.70
11.70
Immunity and Compliance Level
IEC 60601 Test
Level
3 Vrms, 150 kHz to 80
MHz
10 V/m, 80 MHz to 2.7
GHz
CGM MB-1101
80 MHz to 800 MHz
0.04
0.11
0.35
1.11
3.50
Actual Immunity
Compliance Level
Level
3 Vrms, 150 kHz to 80
3 Vrms, 150 kHz to 80
MHz
MHz
10V/m, 80 MHz to 2.7
10 V/m, 80 MHz to 2.7
GHz
GHz
CGM MB -1101
800 MHz to 2.7
GHz
0.07
0.22
0.70
2.21
7.00

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis