Herunterladen Inhalt Inhalt Diese Seite drucken

Viatom Checkme Lite Benutzerhandbuch Seite 16

Vorschau ausblenden Andere Handbücher für Checkme Lite:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 36
When the inputted signal is below the minimum amplitude provided in technical
specifications, erroneous measurements could result.
Portable and mobile communication equipment may affect the performance of this device.
Other devices that have RF transmitter or source may affect this device (e.g. cell phones,
PDAs, and PCs with wireless function).
Guidance and Declaration - Electromagnetic Emissions
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Emission tests
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage Fluctuations /
Flicker Emissions IEC
61000-3-3
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the Health Monitor should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD) IEC
61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
Compliance
Group 1
Class B
Class A
Complies
IEC60601
test
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Electromagnetic environment - guidance
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in
establishments,
establishments and those directly connected to
the public low-voltage power supply
that supplies buildings used for domestic
purposes.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
15
including
domestic
network
Electromagnetic
environment - guidance
Floors should be wood,
concrete
or ceramic tile. If floors
are
covered with synthetic
material,
the relative humidity
should be at
least 30 %.
Mains power quality
should be that of a typical
commercial or hospital
environment.
all

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis