MODE button

There are five manual TENS modes available: (B) Burst, (N)
Normal, (M) / (SD1) / (SD2) Modulated, as well as two EMS
modes; (S) Synchronous, (A) Asynchronous, and the (P)
Program mode. The mode can be changed by pressing the
EN
"MODE" button. In the program
mode P there are 24 pre-defined
programs with fixed settings
availabe.

Setting the parameters

In the manual modes, the
parameters can be selected
by pressing the „SET" button. The keys
and
can be
used to change the values of the selected parameter.

Set frequency

The frequency is adjustable from 1 Hz to 150 Hz. The
current value is shown on the LCD display. The frequency
determines how many electrical impulses are transmitted
through the skin every second. If no instructions regarding
the frequency are given in therapy, set the control to
70 – 120 Hz.