Herunterladen Inhalt Inhalt Diese Seite drucken

Performing A Measurement - Lovibond TB 350 Kurzanleitung

Turbidimeter
Vorschau ausblenden Andere Handbücher für TB 350:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
Description
1 Application Bar
2 Measurement Value
3 Settings
4 Log Button
5 Measurement Button
6 Settings Button
7 Home Button

4.3.2 Performing a Measurement

Please prepare the sample cells included in the scope of delivery before
measurement. Make sure that the sample cells are free of stains, scratches
or blemishes.
1. Collect sample
Ensure sample is representative (well mixed).
Glassware should be free of stains, scratches or other
blemishes.
Rinse collection vessel at least two times with sample
prior to sample collection. 
2. Prepare sample cell
Using a lint free cloth, dry the exterior of the sample
vial.
Use silicone oil to minimize stray light reflections.
3. Insert vial into instrument
Ensure sample is bubble free.
Invert slowly to mix. Do not shake.
4. Close lid
Measurement will not be allowed without lid in place.
5. Check settings
Ensure the measurement settings are correct.
Settings for measurement cannot be changed after
measurement occurs.
EN   TB 350
Function
Navigation path to help the user through the process
steps.
Display of measured sample value with unit type.
Settings applied to measurement are displayed.
Creates a Log when pressed.
Starts a measurement when pressed.
Used to change settings.
Returns user to home screen.
4 Operation
EN
15

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis