Functional description

GB
On
Off
Mode
Sample
MD1X0/200_1b 9b/2020

Operation

Switch the unit on
Selecting a method
ScrollMemory (SM)
To avoid unnecessaryscrollingfor the required test method, the instru-
ment memorizesthe last method usedbefore being switched off. When
the instrument is switched on again, the scroll list comes up with the
last used test method first.
Note:
The description of the methods and reagent handling is not included
in this manual; it is contained in a separatedocument.
Fluoride method
Caution:
A calibration of the method has to be performed before the first use
4
(see
calibration mode Fluoride). The same batch of SPADNSreagent
solution must be usedfor adjustment and test. Theadjustment process
needsto be performed for eachnew batch of SPADNSreagentsolution.

Performing zero

Fill 24 mm vial with 10 ml sample
Close vial
Placesample vial in the sample chamber.
Payattention to the positioning.
33