Herunterladen Inhalt Inhalt Diese Seite drucken

Cleaning Of The Device - Gima SUPER VEGA 1-2 L Bedienungsanleitung

Aspirator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 34
15
Working Condition
Conservation condition and Transport
The technical specifications may change without notice.

CLEANING OF THE DEVICE

Use a soft dry cloth with not – abrasive and not – solvent detergents. To clean the device external parts always
use a cotton cloth dampened with detergent. Don't use abrasive or solvent detergents. Before carrying out
any cleaning and / or maintenance operation, disconnect the appliance from the power supply, unplugging it
or turning off the switch on the device
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE
EQUIPMENT DO NOT GET IN TOUCH WITH LIQUIDS. NEVER CLEAN THE EQUIPMENT WITH WATER.
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glass-
es) to avoid getting in contact with contaminating substances (after each utilization cycle of the machine).
ACCESSORIES SUPPLIES
• Complete aspiration jar 1000ml
• Conical fitting
• Tubes set 6mm x 10mm (trasparent silicon)
• Aspiration probe CH20
• Antibacterial and hydrophobic filter
Available under request with different versions with complete jar 2000ml.
Anti-bacterial and hydrophobic filter: designed for the individual patient to protect patient and machine from
cross-infections. Prevents the liquids, that come into contact with it, from passing through it. Replace it when-
ever you suspect that it may be contaminated and/or it becomes wet or discoloured. Replace the filter every
time it is used if the suction pump is used on patients in unknown pathological situations and where an assess-
ment of indirect contamination is not possible. The filter is not manufactured to be decontaminated, dismantled
and/or sterilised. If, however, the patient's pathology is known and/or there is no risk of indirect contamination,
the filter should be replaced after every work shift or once a month even if the device is not used.
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use.
Reuse may cause cross-infections. Don't use after lapse of the sell-by date
WARNING: Suction tubes for insertion in the human body purchased separately from the machine
should comply with ISO 10993-1 standards on material biocompatibility.
Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and
sterilization. Beyond this limit, the physical-chemical characteristics of the plastic material may show signs of
decay. Therefore, we recommend that you to change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the
said tube. Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable
for reuse. The component must be replaced if there are visible signs of decay of the material constituting the
said component.
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to
the employment of the said component. Therefore, after each cleaning cycle, it is up to the final user to verify
ENGLISH
Room temperature: 5 ÷ 35°C
Room humidity percentage: 10 ÷ 93% RH
Atmospheric pressure: 700 ÷ 1060 hPa
Room temperature: -25 ÷ 70°C
Room humidity percentage: 0 ÷ 93% RH
Atmospheric pressure: 500 ÷ 1060 hPa

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis