Declaration — electromagnetic immunity
This smart blood pressure monitor with ECG & digital stethoscope system is intended for use in the electromagnetic environment specified below.
The customer or the user of the smart blood pressure monitor with ECG & digital stethoscope system should ensure that it is used in such an
environment.
Immunity test
Electrostatic discharge
(ESD)IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge IEC 61000-4-5
IEC 60601 test level
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV,
±15 kV
2 kV for power supply lines
1 kV for input/output lines
1 kV differential mode
2 kV common mode
Compliance level
Electromagnetic environment — guidance
Contact: ±8 kV
Floors should be wood, concrete or ceramic tile. If floors are
Air: ±2 kV, ±4 kV, ±8 kV,
covered with synthetic material, the relative humidity should
±15 kV
be at least 30%.
N/A
The main power quality should be similar to that of a typical
N/A
commercial or hospital environment.
N/A
The main power quality should be similar to that of a typical
N/A
commercial or hospital environment.