Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Sudden
If you disconnect and re-
change in
connect a few minutes later,
sensation
the signal will feel quite a lot
stronger.
The patient is an intended operator. There are no user-serviceable parts inside the
unit, and no calibration is required.
If the above review has failed to resolve your problem, or to report unexpected
operation or events, or to provide feedback, call TensCare or your local dealer
(address on back cover) for advice. Contact TensCare customer service on +44 (0)
1372 723 434. Our staff are trained to assist you with most issues you may have
experienced, without the need to send your product in for service.
European Medical Device Regulation requires that any serious incident that has
occurred in relation to this device should be reported to the manufacturer and the
competent authority in your country. This can be found at:
https://ec.europa.eu/docsroom/documents/36683/attachments/1/translations/en/rendi
tions/pdf

20. GENERAL SPECIFICATION

Waveform
Amplitude
(over 500 Ohm load)
Max intensity
Max pulse energy
Output plug
Channels
Battery
Battery life
Weight
Dimensions
Safety Classification
Environmental Specifications:
Always
disconnecting the lead or the probe.
Asymmetrical rectangular
99.0 mA
+/- 10%
50V zero to peak across 500Ω. Setting 0-99.0 mA in steps
of 00.5 mA
Constant voltage over 470-2000 Ohm
Constant current over 160-470 Ohm
OC cutout below 160 Ohm.
Total output limited to 25 μC per pulse
Fully shielded: touch proof mini USB
Single channel
Rechargeable Li-ion battery
Model: BL-4B
Manufacturer: YJ Power Group Limited
Ratings: 600 mAh, 2.22Wh, 3.7V
At least 15 hours at 50 mA, 300 μs, 50Hz
90g
100 x 65 x 18 mm
Internal power source. Designed for continuous use.
29
return
strength
to
zero
after

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis