Interface commands (general)
Command
Feedback
1)
to pump
LEDx?
LED=0000\r\n
LED=0001\r\n
DSP?
DSP=xxx\r\n
Tax!
-
SPX=xxxx!
-
2)
SMM=xxxx!
-
SDZ=xxxx!
-
2)
SDM=xxxx!
-
2)
SSD=xxxx!
-
-
OK\r\n
-
ERROR\r\n
-
PO/r/n
1) RS 232 interface parameters: 9600 Baud, No Parity, 8 Bit, 1 Stop Bit.
2) for tube-dependent parameters see also section Technical data.
Description
Select parameter: X=0-7
LED off
LED on
LED0?=Volume ml (volume)
LED1?=Flow ml/min (volumetric flow/minute)
LED2?=Tube mm (inner tube diameter)
LED3?=speed rpm
LED4?=rotation
LED5?=rotate clockwise
LED6?=rotate anti-clockwise
LED7?=Pause sec/min/h (pause duration)
Display query: Display current value
Select function key: X=1-7
TA1!=Cal (Calibration)
TA2!=Start/Stop (rotation button)
TA3!=change direction of rotation
TA4!=max (maximum speed)
Ta5!=Up (arrow key)
TA6!=Down (arrow key)
TA7!=Select (select button)
Set pause duration: X=0;2;3
SP0=0001 - 5999 sec./10 e.g. 100 ms
SP2=0010 - 0599 minutes
SP3=0010 - 0750 hours
Volumetric flow ml or μl (tube dependent)
: Xxxx=000.0 - 9999
Setting the speed rpm:
Xxxx=0000 - 9999 (min/max depending on the
gear)
Volume dosing ml or μl (tube dependent):
Xxxx=000.1 - 9999 (not possible when
calibrating)
Adjust inner tube diameter: X=4-digit number
xxxx=0000 (see the following section
"Interface commands: Tube parameters")