Troubleshooting

Troubleshooting
Troubleshooting
Fault
Device does not turn on
Indicators remain off after switching
on the device
Tube movement
Pump does not deliver when rotation
is switched on
Rotation indicator flashes (Hei-FLOW
Expert and Ultimate)
[----]
Display
(Hei-FLOW Ultimate)
Possible cause, remedy
Power supply failure: Check the power
supply connector
Device fuse defective: Contact technical
service
Control/drive/LED defective: Contact
technical service
Tube not properly clamped
Inappropriate tube: Use tube with
adequate wall thickness
No rotation movement, coupling does not
engage: Make sure that the pump head
is in contact with the pump housing
Tube guide system not closed (SP quick
pump head only): Close the tube guide
system
Tube setting wrong (Hei-FLOW Ultimate
only): Verify tube setting
Overheat protection triggered: After
a few minutes and when a certain
temperature level is reached, the pump
switches on again
Power restored after power failure,
restart rotation
Invalid values when calibrating the
dosing volume: Correct values according
to table