Herunterladen Inhalt Inhalt Diese Seite drucken

Specific Arm Module Signals; Specific Base Station Signals - Livassured Nightwatch Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Nightwatch:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

6.2 Specific arm module signals

Continuous green LED
Not connected to the charging adapter: NightWatch alerts are active. A heartbeat
§
is registering, the connection with the base station is in order, the battery is in
order, the wearer is at rest and in a horizontal position.
Connected to the charging adapter: the arm module is fully charged; the battery
§
is in order.
Blinking green LED:
Not connected to the charging adapter: the arm module has detected a heartbeat,
§
but the armband wearer is not at rest yet. The NightWatch alerts are not active
yet. The NightWatch will become active as soon as the wearer is at rest in a
horizontal position for more than two minutes.
Connected to the charging adapter: arm module is still charging.
§

6.3 Specific base station signals

Continuous Orange LED
Base station is connected to power supply.
§
EN
One continuous green LED and one blinking green LED
The arm module is connected to the charging adapter and still charging.
§
Two continuous green LEDs
The arm module is connected to the charging adapter and is fully charged.
§
Blinking green LED:
The arm module has detected a heartbeat, but the armband wearer is not at
§
rest yet.
The heartbeat is being measured, but the NightWatch is not yet fully
active for alerts. As soon as the wearer is at rest in a horizontal position
for more than two minutes, the NightWatch alerts will be fully active.
V3.0
All manuals and user guides at all-guides.com
68

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis