Immunity Test
IEC 60601 Test Level
Electrostatic
± 8 kV contact
Discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply lines
transient/burst
IEC 61000-4-4
± 1 kV for input/output lines
(if applicable)
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
0% U
interruptions and
0% U
voltage variations
70% U
on power supply
0% U
input lines
IEC 61000-4-11
Power frequency
30 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level�
T
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
The VisionPro 26" LED Display and the VisionPro SYNK 26" Wireless LED Display are intended for use in the electromagnetic
environment specified below�
The customer or the user of the display should ensure that it is used in such an environment�
Immunity Test
IEC 60601 Test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2�7 GHz
24
0�5 cycle
T
1 cycle
T
25 cycles
T
5 sec�
T
Compliance Level
3 V
10 V/m
Compliance Level
± 8 kV contact
± 15 kV air
± 2 kV line to ground
± 1 kV line to line
(if applicable)
± 1 kV differential mode
± 2 kV common mode
0% U
0�5 cycle
T
0% U
1 cycle
T
70% U
25 cycles
T
0% U
5 sec�
T
30 A/m
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the displays, including
their cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter�
Recommended Separation Distance:
d = 2√P 80 MHz to 2�7 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m)�
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey (a), should be less than the
compliance level in each frequency range(b)�
Interference may occur in the vicinity of equipment marked
with the following symbol:
Electromagnetic Environment-
Guidance
Floors should be wood, concrete,
or ceramic tile� If floors are covered
with synthetic material, the relative
humidity should be at least 30%�
Mains power quality should be that
of a typical commercial or hospital
environment�
Mains power quality should be that
of a typical commercial or hospital
environment�
Mains power quality should be
that of a typical commercial or
hospital environment� If the user of
the transmitter requires continued
operation during power mains
interruptions, it is recommended
that the Wireless Transmitter be
powered from an uninterruptible
power supply or a battery�
Power-frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment�