Installation

5
Installation
Personnel:
Protective equipment:
If possible, the flow fitting should be installed in vertically running pipe sections
(see
Ä 'Installation Instructions' on page 22) in order to prevent trapped air from
distorting readings.

Installation Instructions

NOTICE!
In the event of extremely high ambient temperatures due to a heat
build-up (> 50 °C) and extreme pipe vibrations (> 150 Hz, 20 m/s2),
we recommend using the wall-mounted version of the LMIT09 with
a separate conductivity value transducer. When using the compact
version, the pipe must be attached both before and after the LMIT09
(see
The measuring bore in the calotte must be aligned parallel to the pipe's
axis, i.e. in the flow direction of the medium.
Deviations in the alignment can lead to distorted readings
(see
417101558 Rev. 06-06.2021
n
Specialist
n
Qualified electrician
n
Mechanic
n
Service personnel
n
Protective gloves
n
Protective eyewear
n
Safety shoes
Ä 'Installation Instructions' on page 22).
Ä 'Installation Instructions' on page 22).
The flow must travel from the bottom up.
The housing of the LMIT09 should ideally be rotated so that the
electrical connections point downwards.
If the transmitter is installed in horizontal pipes, the installation
must be performed from below.