Calibration
enter
cal
Press Cal key 3 seconds and
set Flow Rate calibration
1
The instrument will show the
following screen
Calibration
Wait________0__
4
After checking that the number of
pulses has completely stopped,
press Cal (If the Cal key is
pressed with the number of pulses
set at zero, an error is shown.)
enter
cal
7
The instrument shows the K
Factor calculated.
Calibration
K Factor 2.5
10
0000137936
Before starting the procedure
ensure that:
•There are no air bubbles in the
pipe
•The graduated tank is empty
• The faucet is off
2
The number on the bottom right
indicates the number of pulses
generated by the sensor.
5
Set in the instrument the number
of liters read on the tank
Calibration
10.00 L
8
Press Cal to confirm the new K
Factor
enter
cal
11
rev. 6.1
enter
cal
Press Cal Key
3
Turn on the faucet and fill the
graduated tank up to the required
quantity of water to carry out the
calibration, then turn off the
faucet.
6
press Cal to confirm
enter
cal
9
Normal Status
14