Get a sample water from the
faucet of the probe holder
1
Press_cal
enter
cal
Wait________10s__
4
CLOse flowrate
enter
cal
Press Cal key
7
Wait________100s__
Wait 100 seconds
10
0000137936
Chlorine Probe Calibration
Check the chlorine value by
handheld controller instrument
2
0.8_ppm
The unit flashing a value, set your
chlorine value to check by
handheld controller instrument
(ex. 1.2ppm Free Chlorine)
5
Close the flowrate in the probe
holder
8
0.0___PPM
Press Cal Key
Calibration During 10 Seconds
Wait________10s__
The unit save the parameters
and exit
11
rev. 6.1
Calibration
enter
cal
Press Cal Key 3 Seconds
Set CL calibration
3
1.2___Ppm
enter
cal
Press Enter
Calibration During 10 Seconds
Wait________10s__
The unit save the parameters
6
ARE you sure?
enter
cal
Select yes, if you are sure that the
flowrate is closed and confirm with
Enter key.
9
12