Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration - Pepperl+Fuchs LUC-T10 Bedienungsanleitung

Füllstandmessung mit ultraschall
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

#
VH
Entry
1
V9H5
333
H
2
V8H2
(0 ... 1)
H
3
V0H1
E (m / ft)
H
4
V0H2
F (m / ft)
H
5
V0H3
0 ... 5
H
BD
E
F
Text
Reset instrument
Units of length
0: meters
1: feet
Empty calibration
Full calibration
Application
Full
100 %
Empty
0 %
V0H3: Applications
0: Liquid
including automatic stirrer
suppression
1: Rapid level change
The height changes quickly.
2: Liquid/dome cover including
automatic stirrer suppression.
The instrument is mounted
under a dome cover.
The max. initial echo factor is
entered as standard.
3: Coarse bulk solids
(grain size from 4 mm/0.16 in)
4: Conveyor belt
14.1

Calibration

Reset
A reset causes most of the instrument
settings to return to the factory settings.
The following parameters are not
affected by a reset:
• all linearization parameters
(V2H0 ... V2H3)
• the Tag number (VAH0)
• the m/ft selection (V8H2)
Note V8H2 units of length
• Units of length remain unchanged
after carrying out a reset.
• They may only be entered directly
after a reset. If the units of length are
changed at a later date then all
subsequent entries must be
repeated.
Display:
V0H0: Level in %
V0H8: Distance in meters/feet
V0H9: Level in meters/feet
Caution!
All entries which follow (linearization,
current output, fixed target
suppression) must be in the same units
as those of the calibration.
e
21

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Luc-t20Luc-t30

Inhaltsverzeichnis