Herunterladen Inhalt Inhalt Diese Seite drucken

Lacquering With Lp/W Lacquer; Repairs/Modifications With Modeling Pastes - egger EL 1 Betriebsanleitung

Lichtpolymerisationsgerät
Vorschau ausblenden Andere Handbücher für EL 1:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4.6

Lacquering with LP/W lacquer

For the surface finish the light-curing soft earmolds (made of LP/W) are always lacquered with
LP/W lacquer (REF 307ff). Furthermore, all presently known hard acrylate-based otoplastic
materials can be coated with LP/W lacquer to improve the fitting of the earmold ("non-skid
coating").
Firs the part of the otoplastic that is to be processed always has to be roughened with
Corundum paper, grit 180, (REF 40704).
Then, fix the BTE earmold onto a piece of nylon tube (REF 97200) and also fix the lac-
quering holder. With an ITE earmold, the special Erkogum can be fixed at the opened
battery valve.
Then, clean and degrease the surface carefully by using the provided hand-spray
pump with a surface cleaner (REF 31600ff).
Apply evenly the lacquer and cure it for 4 – 5 minutes in the light-curing unit.
After the polymerization, the otoplastic (but no ITE device with an built-in electronic –
see separate note below!) is post-cured with LP-Cleaner (REF 31300) in the egger light-
curing unit EL 1 for about 4 - 5 minutes. This is necessary to remove the inhibition layer
(sticky surface).
Then, the otoplastic is rinsed under water and blown dry.
ITE earmolds with an built-in electronic can be coated with LP/W lacquer, too.
Please observe, however that the otoplastic should not be put into the LP-
cleaner bath! To remove the inhibition layer you use a surface cleaner moisted
cloth and wipe off the surface thoroughly.
4.7
Modifications/Repairs with Modeling Pastes
The LP/H modeling pastes can be universally used for various modifications and repairs with
all kinds of hard UV-curing polymerizates or acrylates. These are available in the colors trans-
parent, beige, red and blue.
 First the part of the otoplastic that is to be processed always has to be roughened
with Corundum paper, grit 180, (REF 40704).
 Then, clean and degrease the surface of the otoplastic carefully with Surface
Cleaner (REF 31600).
 Directly apply the modeling paste out of the tube with a modeling instrument (REF
31215) and smooth the transitions to the ground.
 Then, depending on the coating thickness applied, let it cure for about 2 – 4 minutes
in the light-curing unit egger EL 1.
 After the polymerization of the surface, wipe off the otoplastic by using a surface
cleaner moistened cloth.
 If required, rework the otoplastic with a cutter and prepare it for lacquering as de-
scribed in 4.5/4.6.
Betriebsanleitung/Operating Manual EL 1 REF 51100/51120V03-2014-12
Seite / Page 21 von / of 24

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

51100

Inhaltsverzeichnis