Herunterladen Inhalt Inhalt Diese Seite drucken

Bosch SCT Serie Prüfbuch Seite 34

Kalibrier- und justiervorrichtungen
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

en
en
34 | SCT / CCT / EL | Checking standard components
!
Maximum permissible observational error in x-direc-
tion due to play: 5.2 mm.
Measure in case of impermissible observational error:
Check the fit clearance of the mounting hole for the
P-Assist S5 and, if necessary, replace the P-Assist
S5.
Check laser mount A on the L -bracket and, if neces-
sary, replace the precision measuring bar.
4.2.4
Checking the P-Assist S5 in laser mount A
1. Set the precision measuring bar to a height of
1000 mm.
2. Using the positioning feet, level the SCT 415 on the
base plate with the aid of the spirit levels on the L -
bracket (Abb. 23, item 2 and item 3).
3. Fasten the P-Assist S5 (Abb. 17, item 1) to laser
mount A on the L -bracket and rotate counterclockwi-
se as far as possible.
4. Switch on the P-Assist S5.
5. Align test board (Abb. 17, item 3) with the 0-point
at a distance of 3000 mm with the aid of the laser
beam.
6. Using double-sided adhesive tape, fasten the test
board (aligned with 0-point) (Abb. 17, item 3) to the
wall at a height of 1000 mm.
7. Remove the P-Assist S5.
8. Fasten the reference laser (Abb. 17, item 1) to laser
mount A on the L -bracket and rotate counterclockwi-
se as far as possible.
 The laser beam of the reference laser (Abb. 17,
item 2) marks the test board.
9. Check the difference (Abb. 17, item 4) between the
0-point on the test board (Abb. 17, item 3) and the
laser beam of the reference laser (Abb. 17, item 2).
|
1 690 386 058
2018-01-30
3
4
2
3
2
2
Abb. 17: Laser beam of the reference laser on the test board
1 P-Assist S5/ reference laser
2 Laser beam of reference laser
3 Test board
4 Difference between 0-point on test board and laser beam of
reference laser
!
Maximum permissible difference between laser
beam of the reference laser and 0-point on the test
board: 7.8 mm
Measure in case of impermissible observational error
Check laser mount A on the L -bracket and, if neces-
sary, replace the precision measuring bar.
Check the play of the P-Assist S5 and, if necessary,
replace the P-Assist S5.
2
1