If the M Acoustic Earhook is exposed to moisture
or humidity, it is best to dry it daily by wiping with
a dry tissue.
If utilizing a dome, it should be replaced every three
months, or when it becomes stiff or brittle.
Replace the earwax protection on the speaker unit
every four to eight weeks, or sooner if it becomes
visibly dirty, blocked, or the sound seems quieter
or lower.
Remove the M Acoustic Earhook before using hair
spray or applying cosmetics, as these products may
damage the M Acoustic Earhook.
To clean your M Acoustic Earhook:
»
Remove the dome from the speaker unit.
»
Using a lint-free cloth or moist cleansing wipe, gently clean
the dome and speaker unit. Never use cleaning agents such
as household detergents or soap to clean any parts of your
M Acoustic Earhook.
»
Prior to reattaching the dome to the speaker unit, ensure that
all parts are completely dry.
Recommended Operating, Storage Temperature, and
Humidity Ranges
C
ondition
Operating Temperature
Storage Temperature
Relative Humidity
M
iniMuM
0˚C (32˚F)
-20˚C (-13˚F)
0%
M
axiMuM
45˚C (115˚F)
55˚C (131˚F)
95%